(discontinued)
Automatic calibration portable water hardness tester kit
Features:
1.It is for measurement of water hardness (the total concentration of calcium ion and magnesium ion).
2. Electrode adopts model 601-F water hardness combination electrode with patent structure. It consists of measurement electrode, reference electrode and temperature probe. Replaceable measuring head, steady potential, fast response, easy to use.
3.Meter adopts intelligent chip design, automatic calibration, automatic temperature compensation, time display, data storage, low voltage display, etc.
4. Ability to switch among eight units: mmol/L, mg/L (CaCO3), mg/L (CaO), mmol/L (Boiler), mg/L (Ca). Three options of three units : ºfH(France),ºdH (Germany)and ºeH(UK)from WH-Link communication software.
5. Equipped with B1,B2 and B3 calibration solution.
6. IP57 waterproof rating, equipped with portable case, calibration solution and other accessories.
Water Hardness
|
Temp. compensation
Calibration Solution
Calibration Mode
|
(0 ~ 10)mmol/L; (0 ~ 401)mg/L(Ca);
(0 ~ 1000)mg/L(CaCO3); (0 ~ 100)ºfH(France Degree);
(0 ~ 561)mg/L(CaO); (0 ~ 56)ºdH(German Degree);
(0 ~ 20)mmol/L(Boiler); (0 ~ 70)ºeH (England Degree).
0.01 and 0.1 water hardness units
+/-5% FS
(5 ~ 50)℃(automatic)
B1 Calibration Solution – 2.00×10-2 mmol/L
B2 Calibration Solution — 2.00×10-1 mmol/L
B3 Calibration Solution – 2.00 mmol/L
(B3Calibration Solution also can be used for the soaking solution of electrode)
(a)B1/B2 Calibration-using B1 and B2 Calibration Solution,suitable for < 2.00×10-2 mmol/L low concentration water quality,for example, boiler
water. (b)Calibration—using B2 and B3 Calibration Solution,suitable forgeneral water quality.
|
Temperature
|
Measuring range
Resolution
Accuracy
|
0 ~ 60℃
0.1℃
+/-0.5℃
|
Other Technical Parameter
|
Data storage
Content Storage
Power Dimension and weight Quality & safecertificate |
99 groups
serial number of measuring value, measuring value, units, temperature, ATC or MTC, date, time and calibration way.
Two AA cells(1.5V×2) Meter:(65×120×31)mm/180g; Carry Case: (255×210×50)mm/790g IISO9001:2000 and CE |
Working conditions
|
Ambient temperature
ambient humidity
|
(5~35)℃
≤85%RH
|
2. Model 601-S water hardness electrode( With two measruing heads and one screw plug ) 1pc
3. AA cell (1.5V ) 2pcs (included)
4.100mL hardness calibration solution ( B1, B2 and B3 ) 1btl/each
5. 20mL test cup 6pcs
6. Test cup pedestal 1pc
7. Carring case 1pc
8. Operation manual 1 set