(discontinued)
Water hardness meter with B1-3 calibration solution
2. Electrode adopts model 601-F water hardness combination electrode with patent structure. It consists of measurement electrode, reference electrode and temperature probe. Replaceable measuring head, steady potential, fast response, easy to use.
3. Meter adopts intelligent chip design, automatic calibration, automatic temperature compensation, time display, data storage RS232 output, etc.
4. Ability to switch among eight units: mmol/L, mg/L (CaCO3), mg/L (CaO), mmol/L (Boiler) and mg/L (Ca). Three options of three units: ºfH(France),ºdH (Germany)and ºeH(UK) from WH-Link communication software.
5. Equipped with B1,B2 and B3 calibration solution.
6. Equipped with WH-Link communication software, connect testing data of meter to computer.
Water Hardness
|
Temp. compensation
Calibration Solution
Calibration Mode
|
(0 ~ 10)mmol/L; (0 ~ 401)mg/L(Ca);
(0 ~ 1000)mg/L(CaCO3); (0 ~ 100)ºfH(France Degree);
(0 ~ 561)mg/L(CaO); (0 ~ 56)ºdH(German Degree);
(0 ~ 20)mmol/L(Boiler); (0 ~ 70)ºeH (England Degree)。
0.01 and 0.1 water hardness units
+/-5% FS
(5 ~ 50)℃(automatic)
B1 Calibration Solution – 2.00×10-2 mmol/L
B2 Calibration Solution – 2.00×10-1 mmol/L
B3 Calibration Solution – 2.00 mmol/L
(B3Calibration Solution also can be used for the soaking solution of electrode)
(a)B1/B2 Calibration-using B1 and B2 Calibration Solution,suitable for < 2.00×10-2 mmol/L low concentration water quality,for example, boiler
water. (b)Calibration—using B2 and B3 Calibration Solution,suitable forgeneral water quality.
|
Temperature
|
Measuring range
Resolution
Accuracy
|
0 ~ 60℃
0.1℃
+/-0.5℃
|
Other Technical Parameter
|
Data storage
Content Storage Communication Connector Power Dimension and weight Quality and safe certificate
|
128 groups
serial number of measuring value, measuring value, units, temperature, ATC or MTC, date, time and calibration way.
RS232
DC9V (160×190×70)mm/750g IISO9001:2000 and CE |
working conditions
|
Ambient temperature
ambient humidity
|
(5~35)℃
≤85%RH
|
2. Model 601-F water hardness electrode(With two measruing heads and one screw plug ) 1pc
3. Model 602 Flexible electrode holder 1pc
4. 250mL water hardness calibration solution (B1, B2 and B3) 1btl/each
5. 20mL test cup 4pcs
6. Test cup pedestal 1pc
7. WH-Link communication software 1 pc
8. RS-232 communication cable 1pc
9. Operation manual 1book